Zeinalian M, Moafi A, Fadaie Nobari R, Heidari K. Introduction of a new index to compare function of screening labs of thalassemia prevention program in Isfahan province . Sci J Iran Blood Transfus Organ 2009; 6 (2) :131-139
URL: http://bloodjournal.ir/article-1-326-en.html
Abstract:   (14994 Views)

  Abstract

 Background and Objectives

 Although major beta-thalassemia prevention and control program is one of the most successful programs in genetics field, there are new cases of major beta-thalassemia due to laboratory errors. So the laboratory quality control (QC) is very important to diagnose carriers in preventive programs. At present, there is not a distinct and measurable index for function assessment and comparison in screening laboratories this urged us to create an objective, measurable and explicit index for health care givers. For this purpose, we compared the function of screening labs in districts of Isfahan Province.

 

 Materials and Methods

 In this cross-sectional study, we collected the data from "pre-marriage screening program" in 18 districts (19 screening labs) covered by Isfahan University of Medical Sciences from 2006 to 2008. The percentage rates of "screened males with low blood indexes" in every 3 months (P index) were assessed then, the average and coefficient variants in duration of 3 years were calculated with excel software. Based on these findings, "screening labs’ functions" were assessed and compared.

 

 Results

 The least "CVs of P index" within 19 labs were found to be in Mobarakeh laboratory (10.4%) and Provincial reference lab (10.9%), respectively. The highest CVs were found in Khour (76.3%) and Tiran and Karvan labs (58.2%), respectively. The P index average in all centers was 14% (13.9% in reference lab) with the highest being in Mobarakeh(24.2%) and the lowest in Khour(8.8%).

 

 Conclusions

 Whereas the number of marriage candidates in each screening lab is relatively stable, the centers with the highest CV value of P index are probably lower in function level as compared with the centers with the lowest CV value of P index (as in reference screening lab). CV of P index may be a good evidence for quality of "screening lab function" and should be evaluated annually. The screening labs with high CV of P index need more attention and evaluation.

 

 Key words : Thalassemia, screening, Iran

 

Keywords: Thalassemia, screening, Iran
Full-Text [PDF 227 kb]   (2590 Downloads)    
Type of Study: Research | Subject: Hematology
Published: 2014/08/16

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2025 CC BY-NC 4.0 | Scientific Journal of Iran Blood Transfus Organ

Designed & Developed by : Yektaweb