Abstract:   (15 Views)
A B S T R A C T
Background and Objectives
Accurate Haemoglobin measurement in thalassemia patients is critical for effective management and prevention of complications. Hemoglobinometers, due to their precision, rapid results, and ease of use, serve as practical tools for monitoring Hb level. This study evaluates the performance of a hemoglobinometer developed by a knowledge-based company, comparing its efficacy with a standard reference device in measuring haemoglobin level.     
Materials and Methods
This cross-sectional study was conducted on 83 thalassemia patients referred to the adult thalassemia clinic. Hb levels were measured using both the test hemoglobinometer and the reference device (Sysmex XP-300). Statistical analysis included the evaluation of sensitivity, specificity, positive predictive value (PPV/NPV)/negative predictive value, Bland-Altman diagram, Intracalss Correlation Coefficient (ICC) and Receiver-Operating Characteristic (ROC) curve was done.
Results
The median hemoglobinometer readings was 9.65 ± 1.40 g/dL with a median of 9.70 g/dL. A strong correlation was observed between the two devices (ICC= 0.96%, 95% CI = 0.94- 0.98). The Bland-Altman diagram revealed no significant difference and the coefficient of determination (R2) was calculated 92.60%. The device demonstrated sensitivity of 83.90%, specificity of 96.30%. PPV of 97.90%, NPV of 74.30%, accuracy of 87.90% and a Kappa coefficient of 74.50%. The optimal cut-off based on the ROC curve, was determined to be 10.30.
Conclusions 
The hemoglobinometer demonstrated high diagnostic accuracy for detecting positive cases (PPV: 97.90%), with only one false positive result among 48 positive readings, indicating its suitability for identifying thalassemia patients requiring blood transfusion. However, among negative cases (NPV: 74.30%), 9 false negative cases were recorded out of 35 readings. It emphasizes the necessity of re-evaluating negative results using a reference device or other diagnostic methods.
 
Full-Text [PDF 842 kb]   (14 Downloads)    
Type of Study: Research | Subject: Hematology

Add your comments about this article : Your username or Email:
CAPTCHA

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2025 CC BY-NC 4.0 | Journal of Iranian Blood Transfusion

Designed & Developed by: Yektaweb